Ac Current Measurement; True Inrush ® Measurement; Continuity Measurement - AEMC 603 User Manual

Clamp-on meter
Table of Contents

Advertisement

4.2.4

AC Current Measurement

Measurement
Range
Specified
Measurement
Range (2)
Accuracy
Resolution
Note (1)
The display indicates "OL" above 3000A in PEAK mode.
- Bandwidth in AC = 1 kHz
Note (2)
In AC, any value between zero and the min. threshold of the
measurement range (0.25A) is forced to show "----" on the display.
Specific Specifications in MAX/MIN mode (from 10Hz to 1kHz, and from
0.30A in AC):
Accuracy (with zero corrected): add 1% R to the values of the table
above.
Capture of the extreme: approximately 100ms.
4.2.5
True Inrush
Measurement Range
Specified Measurement Range
Accuracy
Resolution
Specific Specifications in PEAK mode in True InRush
Accuracy: add ± (1.5% R +0.5A) to the values in the table above.
PEAK capture time: 1ms min. to 1.5ms max.
4.2.6

Continuity Measurement

Measurement Range
Open-circuit Voltage
Measurement Current
Accuracy
Buzzer Triggering Threshold
0.25 to 99.99A
0 to 100% of the measurement range
± (1% R + 10cts)
0.01A
®
Measurement
20 to 2000A
Adjustable from 1 to 999Ω (40Ω is the default)
100.0 to 999.9A
± (1% R +3cts)
0.1A
AC
0 to 100% of the measurement range
± (5% R + 5cts)
0.0 to 999.9Ω
≤ 3.6V
550µA
± (1% R +5cts)
33
1000 to 2000A (1)
1000 to 1500A
± (1.5% R +3cts)
1500 to 2000A
± (2% R +5cts)
1A
20 to 3000A
1A
®
(from 10Hz to 1kHz):