Baud Rate - AEMC PowerPad 3945-B User Manual

3-phase power quality analyzer
Table of Contents

Advertisement

buttons to select the desired measurement range of 200, 100 or 5
Amps. If the 5 Amp range is selected, a choice for programming the ratio
is offered.
-
To change the ratio, press the
Pressing either of the
-
After making this selection, press the
value. Each press of the
to the left or right, accordingly. Use the
position as desired. The primary range can be set between 1 and 2999.
• Press the
menu will once again be displayed on the screen.
In addition to the 4 current probe choices, there is an adapter selection. This
selection allows the operator to use current probes that have a current output
with PowerPad
selection. The primary current can be programmed from 5 to 2999 Amps.
The secondary current can be set to 1 or 5 Amps. Adjustment for the ratio is
performed in the same way as described on the previous page for the MN
probe.
4.1.7

Baud Rate

• Highlight BAUD RATE with the
• Choose from the baud rate values: 2400, 4800, 7200, 9600, 19200, 38400,
57600 or 115200 with the
• Press the
menu will once again be displayed on the screen.
For transfer of data between the PowerPad
speeds must be identical on both sides.
Power Quality Analyzer Model 3945-B
button to apply the current sensor selection. The Configuration
. The ratio for these probes can be programmed with this
®
buttons.
button to apply the current sensor selection. The Configuration
button to highlight the secondary value.
buttons will toggle this value to either 1 or 5.
buttons will move the highlight one digit
buttons, then press the
®
button to adjust the primary
buttons to change each digit
button.
and PC, the communication
35