3.10 AC Current Probe MiniFlex
When installing probes, face the arrow on the probe in the direction of
the load.
Nominal Range: 1000A
Measurement Range: 10A to 1000A
Currents below 10A will be displayed as zero with this probe.
Probe Output Signal: 47µV
NOTE: Output is proportional to the amplitude and frequency of the measured current.
Sensor: Length = 10" (250mm); Ø = 2.75" (70mm)
Safety: EN 61010-1 and 2, Pollution Degree 2, 600V CAT IV, 1000V CAT III
Reference Conditions:
Ambient temperature
Humidity
Position of conductor in the sensor
Continuous magnetic field
External alternative magnetic field
External electric field
Frequency
Type of signal measured
Power Quality Analyzer Model 3945-B
)
AC
(% of the output signal)
-4 to 140°F (-20 to 60°C)
10 to 90% RH
10Hz to 20kHz
Conductor in contact with
AC
/1000A
AC
10 to 100A
≤3% ± 1ct
≤0.5°
Range
Any position
the sensor
Sensor
®
(3945-B accuracy included)
max
AC
at 60Hz
AC
64 to 82°F (18 to 28°C)
<40A/m (earth's magnetic field)
100 to 6500A
≤2%
≤0.5°
Error
0.2% per 10°C
0.5%
0.5%
2%
(4% near latching system)
1%
(2% near latching system)
20 to 75% of RH
Centered
None present
None present
10 to 100Hz
Sinusoidal
25