Minimum And Maximum Current Value Measurements - AEMC PowerPad 3945-B User Manual

3-phase power quality analyzer
Table of Contents

Advertisement

5.1.5

Minimum and Maximum Current Value Measurements

MAX
AVG
MIN
PEAK+
1
PEAK -
RMS
Crest Factor - refreshed every 250ms (but calculated every second).
Select 3V or 4A with
or voltage. Select L1, L2 or L3 to obtain these values for an individual phase.
MIN, AVG and MAX values are measured as soon as the power is turned ON.
Pressing the enter button will reset these values and begin a new update.
The MAX and MIN measurements are calculated every half period (e.g.
every 10ms for a 50Hz signal). The AVG measurements are calculated every
second. However, the MAX, AVG and MIN measurements are refreshed every
250ms.
Power Quality Analyzer Model 3945-B