5.2.3

Power and Direction Flow Analysis

59.98Hz
VAh03
%
100
0
-100
V
Harmonics on this screen are shown with either a positive or negative orientation.
Since the bar selected in this example is negative, it indicates that it is a harmonic
from load to source.
By convention, positive harmonics are from supply to load, and negative har-
monics are from load to supply.
The signs are only available in harmonic power measurement.
The example above shows a power harmonic display with the 3
lighted by the cursor. This harmonic is generated by the load.
Harmonic direction is influenced by probe installation on the proper phase
and arrow on probe facing the load.