Securing to the
countertop
Venting the
circulation pump
Protective foil/
countertop
protector
To improve stability, the machine must be secured to the countertop
after it has been aligned.
Open the door.
Screw the machine to the continuous countertop through the holes
in the front trim on the left and right.
Please contact Miele Service to secure it at the sides to adjacent
cabinetry.
 The gaps between a built-in machine and adjacent cabinetry
must not be sealed, e.g., with silicone sealant, as this could
compromise ventilation of the circulation pump.
The protective foil supplied protects the countertop from damage
caused by steam when the door is opened. It should be positioned
underneath the countertop above the machine door.

Electromagnetic compatibility (EMC)

The machine has been tested for electromagnetic compatibility in
accordance with EN 61326-1 and is suitable for operation in
commercial environments, such as hospitals, medical practices, and
laboratories, as well as other similar environments which are
connected to the domestic power supply.
The machine's high frequency (HF) energy emissions are very low and
are therefore unlikely to interfere with other electronic appliances in
the vicinity.
Flooring in the installation area must be wood, concrete, or tiled.
Synthetic flooring must be able to withstand a relative humidity level
of at least 30% to minimize the risk of electrostatic discharges.
The quality of the power supply should comply with that found in a
typical commercial or hospital environment. Check that the power
supply voltage is within a range of +/-10% of its nominal value.
Installation
155