Light Indicators And Audible Sound - Sirona Orthophos 3 Maintenance Manual

Hide thumbs Also See for Orthophos 3:
Table of Contents

Advertisement

Sirona Dental Systems GmbH
ON
ON
I
O
Light Indicators
Unit ON LED:
Depress the main switch into the "
The unit adjusts itself automatically, wait about 1 minute.
The "Unit ON LED" in the upper left corner of the Multitimer will then indicate that the unit is ON.
Digital displays at the Multitimer:
The program and exposure parameters employed with the last patient appear.
A shows you sequentially the P1, P6 and P11 exposure programs (with the diaphragm 1 set) and
the respective maximum exposure time. Press – + program keys for correct display check.
With diaphragms 3 and 4 the exposure times for tele-exposure light up.
B gives you the kV/mA matched value pair. The LED over the respective patient symbol must light up.
LED over the return key R must blink. (When LED blinks, unit is not ready for the exposure).
C lights up, when Ceph–mode is selected (see ORTHOPHOS 3C Operating Instructions for
Tele-Exposures).
See also Operating Instructions under "Preparing the Exposure" subchapter "Switching ON the Unit".
59 01 603 D3285
D3285.103.04.02.02
, Audible Sound
at the Multitimer
I
" position to turn unit ON.
C
Unit ON LED
C
P
I
A
S
6 8
kV
B
I 0
mA
at the Unit
Maintenance Instructions
Ceph-mode only
X-RAY
Ready LED
Multitimer
.
continues
5