Sirona Orthophos 3 Maintenance Manual page 15

Hide thumbs Also See for Orthophos 3:
Table of Contents

Advertisement

Sirona Dental Systems GmbH
Multitimer
0 0 7
E 2
kV
0 4
mA
1
6
.
2 – 4 – 1
4
.
Memory
5
.
S23
Select service routine S.02 at the Multitimer
4. Press Memory key.
5. Press service key S23 until (approx. 4s) the digital displays disappear.
6. Within 3s press patient symbol keys in the sequence 2 – 4 – 1.
- Service mode is now selected.
If the sequence is not observed when selecting the service routine, if a wrong key is pressed or the
time is exceeded, the system switches automatically to normal mode.
7. Press service key S23.
- Service routine S.0 I appears on the digital display.
8. Press + key to select service routine S.02.
9. Press service key S23.
- On the kV display appears 00.
10. 1. Press + key to select service code 02.
2. Press service key S23 twice
3. Select exposure time 3.20s.
4. Select 80kV.
59 01 603 D3285
D3285.103.04.02.02
X-RAY
mA
2
4
7
X-RAY
S.0 2
0 2
kV
mA
(kV/mA blinking)
X-RAY
S. 0 I
kV
.
S23
10.
1.
(X-ray head and cassette carriage remain stationary):
Maintenance Instructions
X-RAY
8
S. 0 2
0 0
kV
mA
9
.
S23
X-RAY
c
3.2 0
3.
s
8 0
kV
4.
I 0
mA
2. S23 twice
.
15

Advertisement

Table of Contents
loading

This manual is also suitable for:

Orthophos 3 ceph

Table of Contents