Maintenance Instructions
1
PAN
Correct
X
position
of X-ray
beam
X-RAY
c
3.2 0
s
8 0
kV
I 0
mA
Darken the room.
10. Activate X-radiation to check correct position of X-ray beam. CAUTION RADIATION!
If X-ray beam is not in correct position adjust the X-ray beam position to the corresponding mark 1
of the alignment tool.
To do this, remove wheel cover of X-ray head (see Service Manual).
Activate X-radiation. CAUTION RADIATION!
Activate X-radiation only for as long as you need to recognize the X-ray beam position.
X Beam correction 'HIGH – LOW': Loosen 2 screws E by 1 turn. Set beam via screw A.
Retighten screws E.
Y Beam correction 'VERTICAL': Loosen 2 screws E. Set beam via screws B and C.
Retighten screws E.
Z
Beam correction 'RIGHT – LEFT': Loosen 2 screws E. Set beam via screws B and C.
Retighten screws E.
Remove beam alignment tool and switch unit OFF.
NOTE: Final check of X-ray beam position under "Phantom Radiograph".