Sirona Orthophos 3 Maintenance Manual page 9

Hide thumbs Also See for Orthophos 3:
Table of Contents

Advertisement

Sirona Dental Systems GmbH
VDC
8
nominal 7.2 V
7
6
5
4
3
2
1
0
kV – ramp diagram
with program P1 and 72kV/10mA set on the Multitimer.
ATTENTION
The kV - ramp can be changed at the customer's request.
59 01 603 D3285
D3285.103.04.02.02
ramp
1.6s
4.1s±0.4s
5.65s±0.5s
1
2
3
4
5
6
nominal 7.9 V
7.3s±0.6s
11.3s
complete P1
exposure time
7
8
9
10
11
Maintenance Instructions
s
exposure time
9