Single and Dual Input Analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468

5.5.1 Wash Cycle Configuration (applicable only to Alarm 3)

A3: Type
Wash
set to
Cont.
Pulsed
-----
Wash Mode
Hours
Mins
15
Wash Frequency
Mins
Secs
15
Wash Duration
Mins
1. 0
Recovery Period
Config. Alarm 3
Continuous
Pulsed
Fig. 5.1 Pulsed and Continuous Wash Cycles
36
Wash Mode
Select the wash mode required.
Cont.
Pulsed
Wash Frequency
Set the wash frequency required.
Wash frequency is set in 15 minute increments between 15 and 45 minutes, then in 1
hour increments between 1 and 24 hours.
Wash Duration
Set the wash duration required.
Wash duration is set in 15 second increments between 15 and 45 seconds, then in 1
minute increments between 1 and 10 minutes.
Recovery Period
Set the recovery period required, between 0.5 and 5.0 minutes in 0.5 minute
increments.
Option board fitted and analog features enabled (see section 7.3, page 58) - Alarm 4
Config. Alarm 4
configuation is identical to Alarm 1.
Option board not fitted or option board fitted and analog features disabled (see section
CONFIG. OUTPUTS
7.3, page 58) - see section 5.6, page 38.
Wash Duration
1s 1s
-
(continuous) the relay remains energized for the wash duration
-
the relay is switched on and off every second for the duration of
the wash, - see Fig. 5.1.
Frequency
Recovery Period
5 PROGRAMMING
t
t