Statistical Operations on Stored Data

This instrument can perform statistical operations (maximum, minimum, agerage, standard deviation) on
stored data. These operations are performed on all the chroma or color difference data stored in one page.
[Operating Procedure]
Change to the color space for the statistical operations.
Press the Statistical Operation key
1
measurement screen.
• The statistical operation screen appears.
Select a page using the
2
Press the Measure/Enter key
3
• The absolute value/color difference screen appears for the sta-
tistical operation.
Select "ABS. VALUE" or "DIFF. VALUE" using the
4
key .
Press the Measure/Enter key
5
• The results of the statistical operation are displayed after the
operating screen.
[MAX]... maximum
[AVG] ... average
• Pressing the Escape key
110
1
key.
.
.
[MIN] ... minimum
[SD] ... standard deviation
returns to the previous screen.
Esc
while in the
Statistic
PQRS
2,4
3,5
[STATIST.]  1/1 
 P02
 P01
 P00
P00  
C P
 [0030]
Statistical operation screen
[STATIST.]  1/1 
 P02
 P01
 P00
P00  
C P
 [0030]
Statistical operation screen
[STATIST.]
 ABS. VALUE
 DIFF. VALUE
P00  
 [0030]
C P
Absolute value/color
difference screen
   [MAX]  [MIN] 
L*  89.22  86.07
a*   3.13 − 1.74
b*  35.58  15.04
   [AVG]  [SD]  
L*  87.23   1.41
a*   1.04   2.05
b*  23.54   8.75