White Calibration

When using the instrument for long periods of time, the displayed value may change depending on
changes in the environment. Therefore, in order to achieve accurate measurements, we recommend
that white calibration is done regularly using the white calibration plate. White calibration should also
be done before measuring after a long time has passed since the instrument was last used.
White calibration should be done under the same temperature conditions as measurement.
1
[Operating Procedure]
1
Press the Calibration key
• The white calibration screen appears.
[CALIB.]
◎WHITE CALIB.
     Y=  0.00
     x=0.0000
     y=0.0000
 USER CALIB.
P00  
No white calibration
data has been set.
Set the calibration data listed on the back of the white calibration plate cover
2
using the
• The cursor moves automatically to the next position when setting data using the numeric pad.
• Use the
• Go to step 3, if white calibration data has already been set.
[Example]
• Illuminants C
• Illuminants D
86
Color Space key
2
4
while in the measurement screen.
C P
 [0000]
key and the numeric pad.
to move the cursor to make corrections.
Y=93.5 x=0.3114 y=0.3190
Y=93.5 x=0.3140 y=0.3318
65
POWER
4
3
[CALIB.]
◎WHITE CALIB.
     Y= 9 3 .5  
     x=0.3 1 1 4
     y=0.3 1 9 0
 USER CALIB.
P00  
White calibration data
has been set.
Ready lamp
White calibration plate
C P
 [0000]