Box tolerance

This judges whether or not the value of the measured color difference is within the range of the box
tolerance limit set for the color difference target color. When setting a box-type tolerance, color differ-
ence tolorance (±) for each color difference target color can also be set.
Set a color difference target color ahead of time before setting a box tolerance.
5,7
Selecting the limit value in P.115 displays the [limit value] selection screen.
[Operating Procedure]
1
Use the
key to select a color difference target
color.
2
Press the Measure/Enter key
• The tolerance type screen appears.
3
Select the "BOX" as the tolerance type.
Press the Measure/Enter key
4
• The limit value setting screen will appear.
• If a color difference tolerance is already set, go to step 6 as the
set color difference tolerance data is displayed.
.
.
1,3,5,7
2,4,6,8
[TOLERANCE]  1/1 
◎TARGET
 T03
 T02
 T01
P00  
C P
 [0001]
Selection screen
[TOLERANCE]
◎TOLERANCE TYPE
 ELLIPTICAL
 BOX
  ΔE*
 BOX & ΔE*
P00   C P  [0001]
Tolerance type screen
[TOLERANCE]
◎TOLERANCE
 A      B
ΔL*
Δa*
 + 0.00 + 0.00
 − 0.00 − 0.00
 C      D
ΔE*
Δb*
 + 0.00    .   
 − 0.00
P00   C   [0001]
P
Setting screen