User Calibration
3
Press the Measure/Enter key
• The multi-calibration setting screen appears.
Item
ALL CHANNELS:
WHITE CAL. TILE: When user calibration is not performed.
Ch**:
Move the cursor using the
4
calibration channel.
5
Press the Measure/Enter key
• Once manual select calibration settings are complete, the display returns to the measurement
screen.
Since settings for calibration channel made in manual select calibration are kept even if the
power is turned OFF (O), it is important to check the calibration channel following steps 1 to
3 above in the following situations.
• When changing the calibration channel settings to measure a different color
• When the instrument has been left unused for long periods of time
138
.
Multi-calibration is set.
(See P.86 "White Calibration")
Manual select calibration is set
(See P.137 "Setting the Mannul Select
Calibration")
key to and select a
.
[MALTI CAL.] 1/1
 ALL CHANNELS
 WHITE CAL. TILE
 Ch02
 Ch01
P00  
C P
 [0001]
Selection screen
[MALTI CAL.] 1/1
 ALL CHANNELS
 WHITE CAL. TILE
 Ch02
 Ch01
P00  
C P
 [0001]