Error Messages - Minolta CR-400 Instruction Manual

Chroma meter
Table of Contents

Advertisement

Error Messages

The keys will not function if pressed when the error message is displayed.
Message
ER00
Communication performed with inappropriate
command.
COMMUNICATION
• Cable unplugged during communication
ERROR
• Message garbled by external noise, etc.
ER01
[Measuring Head]
1. Number of measurement data exceeds 1000.
MEMORY FULL
2. Number of target color data exceeds 100.
[Data Processor]
3. Number of measurement data exceeds 2000.
4. Number of pages exceeds 100.
5. Number of target color data exceeds 100.
6. Number of user calibration channels exceeds
19.
7. There is insufficient room to transfer all data
to the data processor when importing mea-
surement data.
ER02
Battery voltage drop
LOW BATTERY
ER05
Xenon lamp does not flash properly.
• Xenon lamp has malfunctioned.
ERROR
• Flash circuit has malfunctioned.
IN FLASHING
• Faulty sensor
ER06
Because of the excessively high reflectivity of
the measurement specimen, the operation of
OVERFLOW
measurement data exceeds the performance of
the instrument.
When Y is higher than 160:
ER07
White calibration has not been performed.
WHITE
CALIBRATION
INCOMPLETED
ER08
An overrun or other type of error occurred dur-
ing communication.
COMMUNICATION
• Cable unplugged during communication
ERROR
• Calibration, target color, or option settings
changed on the data processor without hook-
ing it up to the measuring head.
ER10
The data needed for that function to work is not
found.
NO DATA
• Attempting to print or delete nonexistent data
IN MEMORY
• Statistical operations or list displays in pages
without data
146
Symptom/Possible Cause
Corrective Acrion
• Plug in cable with the power OFF.
• Use a lower baud rate if noise is the problem.
1. Connect to data processor and transfer data
there to free up some space, or turn OFF the
data protection setting.
2. Overwrite unnecessary target colors when
registering. Or create more space by connect-
ing to the Data Processor and deleting any
unnecessary target colors.
3,4. If you turn OFF data protect, the oldest
pages will automatically be updated. Or you
can delete some data to free up some space.
5. Overwrite onto unneeded target colors or de-
lete some to free up space.
6. Overwrite onto unneeded channels or delete
some to free up space.
7. Delete some pages or data to free up space.
• Turn OFF the power and replace batteries or
use the AC adapter.
• Repeat measurement.
• If the problem persists, contact your nearest
Minolta-authorized service facility.
• Redo white calibration and the measurement.
• If the problem persists, contact your nearest
Minolta-authorized service facility.
• Perform white calibration.
• Plug in cable with the power OFF.
• Change calibration, target color, or option
settings using the Measuring Head.
• Store some data before attempting

Hide quick links: