Installation Manual
MMI-20058013
Procedure
1. Ground the sensor according to the instructions in the sensor documentation.
2. Ground the transmitter according to applicable local standards, using the
2.5 
Rotating the transmitter on the sensor
(optional)
For easier access to the user interface or the wiring terminals, the transmitter can be
rotated on the sensor in 45° increments, for eight different orientations.
Figure 2-8: Rotating the transmitter on the sensor
A. Clamping ring
Procedure
1. Remove the metal clamping ring from the base of the feed through (refer to
2. Gently lift the transmitter on the feed through until it disengages from the notches
3. Rotate the transmitter to the desired position.
Installation Manual
transmitter's internal or external ground screw.
The earth ground terminal is located inside the power wiring compartment.
The external ground screw is located on the side of the transmitter located below
the transmitter tag.
2-8).
on the feed through. You will not be able to remove the transmitter completely.
Mounting and sensor wiring
April 2023
Figure
17