9.5

Message list

The message list is an overview of all messages that the SIWAREX module can generate. A
message can be quickly identified by the message code (number).
9.5.1

System status message list

Operating errors (code 1000 to 1999)
sorted by code no.
1000 operating error exists
1001 Watchdog
1002 RAM error
1003 Checksum incorrect parameter
1004 Checksum incorrect program
1006 logbook error
1102 ADU error
1104 Undervoltage
1105 Overload
1106 Underload
1107 Legal trade display failure
SIWAREX WP241
Manual, 04/2014, A5E33713528A
Error code
Description and remedy
1000
Group message, at least one operating error exists.
1001
Watchdog, error is displayed for at least 10 seconds. A
serious error has occurred in the function of SIWAREX, e.g.
program error, severe electromagnetic influence on device,
etc.
Contact the SIWAREX Support if the error occurs multiple
times.
1002
RAM error. An error has occurred in the memory; the memory
content is no longer correct. The module must be switched
off. If the error occurs again, SIWAREX is defective.
1003
Checksum error at parameter. Critical error because the
parameters are no longer safe.
1004
Checksum error program code. Critical error because the
program is no longer safe.
1006
Error when writing/deleting,or logbook full
1102
AD converter error when reading in the measured value. If
the error occurs again, make sure that the EMC
recommendations are observed (chapter EMC-compliant
setup (Page 21)).
1104
Undervoltage at sensor cables
1105
Overload of scale (ca. 110%)
1106
Underload of scale (ca. -10%)
1107
The SecureDisplay legal trade display no longer
communicates with the module
Messages
9.5 Message list
141