CombiFlash® R
Installation Guide
f
21. Optional External Gas Port – An inlet for an internal
pump which supplies pressurized air or user-supplied gas
for solvent level sensing, waste container full detection,
and system/column purges. The external gas should be
regulated to 2–5 psig before connecting the source to this
port.
22. Mains Power – Connects the system to AC line voltage.
23. Waste Ports – From left to right:
• Waste Level Sense – The system uses air or
user-supplied gas to measure the hydrostatic
pressure of the waste above the level sense line end.
The system will suspend operation when this level is
too high.
• Diverter Valve Waste – The diverter valve on the
fraction collector arm sends fluid to this waste port.
• Inject Valve Waste – The injection valve is cleaned
after each run. The system directs fluids from this
automatic cleaning routine to this port.
24. Lower Drain Tube – This tube carries away liquids
spilled onto the rack area.

1.6 Safety

Before installing, operating, or maintaining this equipment, it
is imperative that all hazards and preventive measures are
fully understood. While specific hazards may vary according to
location and application, take heed in the following general
warnings:
Liquids associated with this instrument may be classified as
carcinogenic, biohazard, flammable, or radioactive. Should
these liquids be used, it is highly recommended that this
application be accomplished in an isolated environment
designed for these types of materials in accordance with
federal, state, and local regulatory laws, and in compliance
with your company's chemical/hygiene plan in the event of a
spill.
1-8
WARNING