7 Quick start
7.2
Flowcharts for analysis and recalibration
7.2.1
Start and sequence of a check measurement in spark mode
►
32 Hitachi High-Tech Analytical Science
Start a check measurement
Set argon supply to 3–5 bar
Connect power supply
Switch on computer
Wait until WINDOWS™ has
completely started up
Switch on spectrometer
Start spectrometer software
Select "Analysis" and subprogram
Select "Analysis" menu item;
click "Argon purging" button;
purge for 1 minute, click button
again to complete purging
Measure Control Sample
(see software manual)
Check result
OK
Continue with "Analysing"
chapter 7.2.4, page 35
chapter 5.4, page 19
not OK
Recalibration: Select recalibration
method "Recalibrate Program" or
"Recalibrate Matrix"
chapter 7.2.2, page 33
or
chapter 7.2.3, page 34