3.2.2 AUTO UPLOAD USING THE CONNECTION CABLE

The steps below will guide you through Auto Upload Calibration:
1.
Check to see that the Receiver and OneTouch Ultra meter have the same date and
time.
2.
Take 2 fingersticks with your OneTouch Ultra meter.
3.
Make sure the OneTouch Ultra Meter display is off (screen is blank) before connecting.
To turn the display off, press the M button once.
4.
Using the connection cable provided follow the steps below:
a.
First plug the cable into the connection port of the BG meter. (The connection
port is at the bottom of the meter).
b.
Second connect the wide end of the cable into the Receiver. Make sure the raised
Push button on the cable is facing down. (You must do these steps in order).
5.
After a few seconds, "PC" will appear on the OneTouch Ultra meter screen (shown
below):
6.
The Receiver will display the Connection Screen below. The following screen indicates
that BG readings from the meter are successfully being uploaded to your Receiver for
calibration.
M
C
Connection Screen
Seven
System User's Guide | 51