Guidance and Manufacturer's Declaration- Electromagnetic Immunity
The Transmitter is intended for use in the electromagnetic environment specified below.
The customer or the user of the Transmitter should assure that it is used in such an
environment.
Immunity Test
Electromagnetic
Environment Guidance
Electrostatic Discharge
(ESD)
IEC 61000-4-2
Electrical Fast
Transient /burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage Dips, Short
Interruptions and Voltage
Variations on Power
Supply Input Lines
IEC 61000-4-11
Power Frequency (50/60
Hz) Magnetic Field
IEC 61000-4-8
Note: U
is the a.c. mains voltage prior to application of the test level
T
130 | SEVEN
CONTINUOUS GLUCOSE MONITORING SYSTEM
IEC 60601
Test Level
± 6 kV Contact
± 8 kV Air
± 2 kV for power supply lines
± 1 kV for input/output lines
± 1 kV differential mode
± 2 kV common mode
< 5 % UT (>95 % dip in UT) for
0.5 cycle
40 % UT (60 % dip in UT) for
5 cycles
70 % UT (30 % dip in UT) for
25 cycles
< 5% UT (>95% dip in UT) for
5 sec
3 A/m
Seven Transmitter
Compliance Level
Floors should be wood, concrete
or ceramic tile. If floors are
± 6 kV Contact
covered with synthetic material,
± 8 kV Air
the relative humidity should be
at least 30 %
Not Applicable-
Battery Operated
Not Applicable-
Battery Operated
Not Applicable-
Battery Operated
Power Frequency magnetic fields
should be at levels characteristic
3 A/m
of a typical location in a
typical commercial of hospital
environment
Electromagnetic
Environment
Guidance