Erasing All Records; Error Codes - AEMC 6550 User Manual

10kv and 15kv megohmmeters
Hide thumbs Also See for 6550:
Table of Contents

Advertisement

5.3.2 Erasing All Records

Press the CONFIG button. The instrument requests confirmation of the deletion.
Select OK to confirm or CANCEL to cancel.
67
The instrument in this case completely reformats the memory, which takes a few minutes. During this time, it
displays "WAIT".
The instrument then returns to the top level memory screen. Since there is no longer any record, it displays:

5.4 Error Codes

If an anomaly is detected when the instrument is turned on or in operation, the display indicates a one- or two-
digit number error code. This number identifies the anomaly and states what to do to put the instrument back
into service.
Errors 1 to 9 concern internal circuit board problems and require a repair by a qualified individual.
Error codes 20 to 25 help the repair personnel locate a malfunction
If the recorded data is corrupted, the only way possible to reuse the memory is to erase it completely
(see § 5.3.2). The instrument reports this problem by displaying "CLEAR MEMORY".
All other errors require returning the instrument for repair (see § 8.5).
48
Shop for AEMC products online at:
01 08
CONFIG
ERROR
No data records!
ShopAEMC
www.
MEMORY
! WARNING !
All data sets
will be cleared !
O.K.
CANCEL
Megohmmeter Models 6550/6555
.ca
1.877.766.5412

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

6555

Table of Contents