4.6.1 DAR/PI Measurement

There are several ways of measuring the DAR and the PI:
■ Manual Configuration:
Press the START/STOP button. Wait one minute for the DAR or ten minutes for the PI (if the default values are
used).
Press the START/STOP button again to stop the measurement.
■ Automatic Configuration (preferred)
Press the CONFIG button.
38
Shop for AEMC products online at:
START/STOP
START/STOP
Total Run Time
CONFIG
www.
499 MΩ
502 V
Set: 500 V
Elapsed Time 00:10:08
kΩ
MΩ
1
1
10 100
10
100
502 MΩ
502 V
Elapsed Time 00:10:10
DAR (30s/60s)
PI (1.0m/10m)
Capacitance
CONFIG
Manual Stop
Manual Stop + DD
Timed Run (m:s)
Timed Run + DD
DAR (s/s)
PI (m/m)
ShopAEMC
.ca
BURN
978 nA
GΩ
TΩ
1
10
100
10
BURN
978 nA
2.64
1.05
320 nF
---
2:00
30/60
1.0/10
Megohmmeter Models 6550/6555
1.877.766.5412