2.4.3 CONFIG Button

2.4.3.1 Before the Measurement

If the V-FIXED or V-VAR. test voltages have been chosen, there are two configuration screens. There is only one
for the V-RAMP and V-STEP test voltages.
Press the CONFIG button (press CONFIG again to exit):
Manual Stop: measurements are stopped manually.
Manual Stop + DD: measurements are stopped manually and dielectric discharge ratio (DD) is
calculated at the end of the programmed duration.
Timed Run + DD: measurement is automatically stopped at the end of the programmed duration and
dielectric discharge ratio (DD) is calculated.
DAR: measurement is automatically stopped at the end of one minute
(or of the programmed time, if different).
PI: measurement is automatically stopped at the end of 10 minutes
(or of the programmed time, if different).
It is always possible to stop a measurement during a test with programmed duration by pressing the START/
STOP button.
The ▲▼ arrow buttons are used to select the measurement
configuration.
When Timed Run (test with programmed duration) or Timed Run + DD is selected, the duration of the measure-
ment (m:s) can be set.
Total Run Time
Manual Stop
Manual Stop + DD
Timed Run
Timed Run + DD
DAR (s/s)
PI (m/m)
Megohmmeter Models 6550/6555
Shop for AEMC products online at:
Total Run Time
Manual Stop
Manual Stop + DD
Timed Run (m:s)
Timed Run + DD
DAR (s/s)
PI (m/m)
CONFIG
CONFIG
00:02:00
(m:s)
2:00
30/60
1.0/10
ShopAEMC
www.
CONFIG
---
2:00
30/60
1.0/10
To do this, use the ◄ ►▲▼arrow buttons.
.ca
1.877.766.5412
CONFIG
11