3.4.5 Adjustment of the Test Voltages

Always on the third SET-UP screen.
Adjustable Voltage 1, 2, and 3: Set the values of the 3 adjustable test voltages.
The adjustment range is from 40 to 15,000V.
Press the DISPLAY button to see the following:
Set Step Function 1, 2, and 3: In the case of a measurement with a stepped voltage, used to set the voltages
and the durations of the steps.
Pressing the ► button opens the following screen. Press the DISPLAY button to view the rest of the steps.
Step & Ramp Functions
Step Function 1
Step
Voltage
1
50 V
2
100 V
3
150 V
4
200 V
5
250 V
Total Run Time (m:s)
You can then set the voltage and duration of each of the 10 steps. The total duration of the measurement (Total
Run Time) is calculated by the instrument.
The adjustment range of the voltages is from 40 to 15,000V.
The duration of the steps ranges from zero or 0:01 to 99:59. If a duration is set to zero, the time displayed is -:--.
A duration below 30 seconds is not very useful because acquiring a stable resistance result requires some time.
If either a step voltage or a step duration is set to zero, the step will be set to zero as a whole and will be skipped
during the test.
24
Shop for AEMC products online at:
Step & Ramp Functions
Set Step Function 1
Set Step Function 2
Set Step Function 3
Set Ramp Function 1
Set Ramp Function 2
Set Ramp Function 3
DISPLAY
Duration (m:s)
0:10
0:10
0:10
0:10
0:10
DISPLAY
1:40
ShopAEMC
www.
Step & Ramp Functions
Step Function 1
Step
Voltage
Duration (m:s)
6
300 V
7
350 V
8
400 V
9
450 V
10
500 V
Total Run Time (m:s)
Megohmmeter Models 6550/6555
.ca
1.877.766.5412
0:10
0:10
0:10
0:10
0:10
1:40