4.4.8 During the Insulation Measurement
Press the START/STOP button to start the measurement.
The instrument generates high voltage. To indicate that the measurement is in progress, the instrument emits
an audible beep every ten seconds (if the buzzer is switched on) and the START/STOP button lights up red.
If the test voltage generated is >5000V, the START/STOP button blinks.
After a few seconds, the measurement is displayed in digital form and in analog form on a bargraph.
If the measurement is unstable, it is possible to apply a digital filter by pressing the FILTER button.
It is possible to view the available result values by pressing the DISPLAY button.