Advertisement

5 Operation
5.5 Main home screen
40
5.5.1

Self Test

After booting the SIROLaser Blue will automatically perform a self test
including a status check of the following matters:
● Wireless foot control vs. finger switch
Self Test screen with foot control
● USB port
In addition, you will be informed when the next calibration check or when
the next servicing is due.
The statuses are displayed on the home screen.
Foot control
Self Test of the foot control
If the wireless foot control is selected in the settings the self test may
result in:
The wireless foot control is selected and the appropriate symbol for foot
control will be displayed on the home screen confirming selected.
If the wireless foot control is not selected the appropriate symbol for foot
control will not be displayed on the home screen.
NOTICE
If the wireless foot control is selected but not detected, please perform
the registry of the wireless foot control and/or check the battery, see
chapter "Install wireless foot control – optional [ → 34]".
If the wireless foot control remains not selectable please contact your
local dental dealer or an authorized Customer Service Department for
technical support.
Finger switch
If the finger switch is selected in settings the self test may result into:
With Foot control
The finger switch is connected and the appropriate symbol for finger
switch will be displayed on the home screen confirming checked and
selected.
If the finger switch is not selected the appropriate symbol for finger switch
will not be displayed on the home screen.
NOTICE
If the finger switch is defective/missing the unit shows an error message
after booting. In this case please check the cable connection to the
SIROLaser Blue control unit, see chapter "Troubleshooting of simple
defects". If the finger switch remains defective/missing, please contact
your local dental dealer or an authorized Customer Service Department
for technical support.
In general: If any switch is defective laser will be blocked.

Sirona Dental Systems GmbH

Operating Instructions SIROLaser Blue
65 40 863 D3648
D3648.201.01.02.02 10.2015