Sample Collection, Storage, and Preservation
Accuracy Check
Sulfate_UniCell_Turb_Eng_Ody.fm
Analyze samples within 3 hours after collection for best results. Samples may be
stored up to 28 days at 4 °C (39 °F). Warm samples to room temperature before
analysis.
Standard Additions Method (Sample Spike)
1. After reading test results, leave the sample cell (unspiked sample) in the
instrument.
2. Touch
. Touch
Options
additions procedure will appear.
3. Touch
to accept the default values for standard concentration, sample
OK
volume, and spike volumes. Touch
are accepted, the unspiked sample reading will appear in the top row.
4. Prepare three sample spikes. Fill three mixing cylinders with 25 mL of
sample. Use a pipet to add 0.2 mL, 0.4 mL, and 0.6 mL of 2500-mg/L SO
standard, respectively, to each sample and mix thoroughly.
5. Transfer 5 mL of each sample spike to a sample vial and analyze each sample
spike as described in the procedure above, starting with the 0.2 mL sample
spike. Accept each standard additions reading by touching
addition should reflect approximately 100% recovery.
6. After completing the sequence, touch
the standard additions data points, accounting for matrix interferences. Touch
, then select
View: Fit
the sample spikes and the "Ideal Line" of 100% recovery.
See Section 3.2.2 Standard Additions on page 32 for more information.
Standard Solutions
Prepare a 500-mg/L sulfate standard solution as follows:
1. Using Class A glassware, pipet 20 mL of Sulfate Standard Solution,
2500-mg/L, into a 100-mL volumetric flask. Dilute to the mark with
deionized water. Prepare this solution daily. Perform the SulfaVer procedure
as described above.
2. To adjust the calibration curve using the reading obtained with the
500-mg/L standard solution, touch
Touch
Standard Adjust
3. Touch
. Press
On
OK
concentration is used, touch the number in the
actual concentration. Touch
See Section 3.2.4 Adjusting the Standard Curve on page 40 for more information.
. A summary of the standard
Standard Additions
to change these values. After values
Edit
Graph
and touch
Ideal Line
OK
Options
.
to accept the displayed concentration. If an alternate
.
OK
Sulfate
. Each
Read
to view the best-fit line through
to view the relationship between
on the current program menu.
field and enter the
Adjust to
Page 3 of 4
2–
4
Sulfate