Nickel; Dimethylglyoxime Method; Unicell™ Vials - Hach DR2400 Manual

Hide thumbs Also See for DR2400:
Table of Contents

Advertisement

DR/2400
UniCell™ Vials
Scope and Application:
Tips and Techniques
• Undissolved or complexed nickel can only be determined after digestion with the Metal Prep Set (HCT 200).
• Adjust pH of preserved samples to between pH 3–8 before analysis.
• The outside of the vial must be clean before placing it in the instrument. Wipe with a damp towel, followed by a dry one.
• For a proof of accuracy, use the standard solution or standard addition method. (See Accuracy Check).
• See the Instrument Procedure Manual for instructions to adjust the calibration curve, if needed.
• The temperature of the water sample and the sample vial should be between 15–25 °C (59–77 °F).
• Underrange appears on the display when the instrument is zeroed or when the determined sample concentration is below the
operating range listed for this method.
UniCell Vials
Hach Programs
1.
Touch
.
Hach Programs
Select program
.
328 Nickel, HCT 167
Touch
.
Start
Nickel_UniCell_Other_DIMETH_Eng_Ody.fm
For water, wastewater, raw water, and process control
2.
Pipet 4.0 mL of
sample into a sample
vial.

Dimethylglyoxime Method

3.
Pipet 0.4 mL of
Dimethylglyoxime
Solution A (HCT 167A)
into the sample vial.

Nickel

(0.10 to 6.00 mg/L)
4.
Cap and invert the
vial several times to mix.
Nickel
Page 1 of 4