17.
Install the 16-mm
adapter.
Note: See Section 2.6 in the
Instrument Manual for
installation details.
Wipe the reagent blank
and place it into the
adapter.
Blanks for Colorimetric Measurement
Interferences
NitrogenTot_TNT_High_PSD_Eng_Ody.fm
Zero
18.
Touch
.
Zero
The display will show:
0 mg/L N
The reagent blank may be used repeatedly for measurements using the same lots
of reagents. Store the reagent blank in the dark at room temperature (18–25 °C)
for a maximum of seven days. If a small amount of white floc appears prior to
the end of one week, discard the reagent blank and prepare a new one.
The substances in Table 1 have been tested and found not to interfere up to the
indicated levels (in mg/L). Interfering substances that resulted in a
concentration change of ±10% appear in Table 2.
Table 1 Non-interfering Substances
Substance
Barium
Calcium
3+
Chromium (
)
Iron
Lead
Magnesium
Organic Carbon
pH
Phosphorus
Silica
Silver
Tin
19.
Wipe the reagent vial
and place it into the
adapter.
Maximum Level Tested
10.4 mg/L
1200 mg/L
2 mg/L
8 mg/L
26.4 µg/L
2000 mg/L
600 mg/L
13 pH units
400 mg/L
600 mg/L
3.6 mg/L
6 mg/L
Nitrogen, Total
Read
20.
Touch
Read.
Results will appear in
mg/L N.
Nitrogen, Total
Page 3 of 6