Hach DR2400 Manual page 638

Hide thumbs Also See for DR2400:
Table of Contents

Advertisement

Phosphorus, Reactive
Sampling and Storage
Treating Analysis Labware
Cleaning the Pour-Thru Cell
Accuracy Check
Phosphorus, Reactive
Page 4 of 6
Collect samples in clean plastic or glass bottles that have been cleaned with
1:1 Hydrochloric Acid Solution (Cat. No. 884-49) and rinsed with deionized
water. Do not use detergents that contain phosphate for cleaning labware.
Analyze samples immediately for best results. If prompt analysis is not possible,
preserve samples by filtering immediately and storing the sample at 4 °C (39 °F)
or below for up to 48 hours. Warm to room temperature before analyzing.
Clean containers by normal means (do not use detergents containing
phosphorus), then rinse with deionized water. Soak for several minutes in a
1:25 dilution of Molybdovanadate Reagent in deionized water. Rinse well with
deionized water. Dedicate these containers for HR PO
Pour-Thru Cell with this same mixture of Molybdovanadate reagent and
deionized water, and let stand for several minutes. Rinse with 50 mL of
deionized water.
The Pour-Thru Cell may accumulate a buildup of colored products, especially if
the reacted solutions are allowed to stand in the cell for long periods after
measurement. Remove the color by rinsing with a 1:5 dilution of ammonium
hydroxide (Cat. No. 106-49), followed by several deionized water rinses. Invert a
beaker over the glass funnel of the Pour-Thru Cell when not in use.
Standard Additions Method (Sample Spike)
1. After reading test results, leave the sample cell (unspiked sample) in the
instrument. Verify the chemical form.
2. Touch
. Touch
Options
additions procedure will appear.
3. Touch
to accept the default values for standard concentration, sample
OK
volume, and spike volumes. Touch
are accepted, the unspiked sample reading will appear in the top row. See
Standard Additions in the instrument manual for more information.
4. Snap the neck off a Phosphate Voluette
3–
500-mg/L as PO
.
4
5. Prepare three sample spikes. Fill three mixing cylinders (Cat. No. 1896-40)
with 25 mL of sample. Use the TenSette Pipet to add 0.1 mL, 0.2 mL, and
0.3 mL of standard, respectively, to each sample and mix thoroughly.
6. Analyze each sample spike as described in the procedure above, starting with
the 0.1 mL sample spike. Accept each standard additions reading by touching
. Each addition should reflect approximately 100% recovery. Each 0.1-mL
Read
addition of standard should cause an increase of 2.0 mg/L PO
L P.
7. After completing the sequence, touch
the standard additions data points, accounting for matrix interferences. Touch
, then select
View: Fit
the sample spikes and the "Ideal Line" of 100% recovery.
See Section 3.2.2 Standard Additions on page 32 for more information.
. A summary of the standard
Standard Additions
to change these values. After values
Edit
®
Ampule Standard Solution,
Graph
and touch
Ideal Line
OK
3–
analysis. Fill the
4
3–
or 0.65 mg/
4
to view the best-fit line through
to view the relationship between
PhosphorusReac_PT_High_MOL_Eng_Ody.fm

Advertisement

Table of Contents
loading

Table of Contents