Oxygen, Dissolved; Hr (0.3 To 40.0 Mg/L) - Hach DR2400 Manual

Hide thumbs Also See for DR2400:
Table of Contents

Advertisement

DR/2400
Method 8166
®
AccuVac
Ampuls
Scope and Application:
Tips and Techniques
• Analyze samples on-site. Do not store for later analysis.
AccuVac Ampul
Hach Programs
Hach Programs
1.
Touch
.
Hach Programs
Select program
445 Oxygen Di. HR AV
Touch
.
Start
5.
Hold the ampule with
the tip pointing down
and immediately place
the ampule into the
ampule cap.
The cap prevents
contamination from
atmospheric oxygen.
OxygenDis_AccuVac_High_HRD_Eng_Ody.fm
For water and wastewater
2.
Fill a round sample
cell (the blank) with at
least 10-mL of sample.
.
6.
Shake the ampule for
30 seconds.
A small amount of
undissolved reagent will
not affect results.

Oxygen, Dissolved

HR (0.3 to 15.0 mg/L O
3.
Fill a blue ampule cap
with sample.
7.
Touch the timer icon.
Touch
.
OK
A two-minute reaction
period will begin. This
enables the oxygen that
was degassed during
aspiration to redissolve
and react.
HRDO Method
Method 8166
4.
Fill a High Range
Dissolved Oxygen
AccuVac Ampul with
sample. Keep the tip
immersed while the
ampule fills completely.
8.
When the timer
beeps, shake the ampule
for 30 seconds.
Oxygen, Dissolved
Page 1 of 4
)