Hach DR2400 Manual page 341

Hide thumbs Also See for DR2400:
Table of Contents

Advertisement

Method Performance
Summary of Method
Required Reagents
Description
®
FerroVer
Iron Reagent Powder Pillows (for 10-mL sample)............ 1 pillow........... 100/pkg.............21057-69
or
®
FerroVer
Iron Reagent AccuVac
Required Apparatus
Sample Cells, 10-mL, w/cap................................................................... 2 ....................... 6/pkg.................24276-06
Beaker, 50-mL ............................................................................................ 1 ....................... each.................... 500-41H
Required Standards
Iron Standard Solution, 100-mg/L................................................................................... 100 mL...............14175-42
Iron Standard Solution, 10-mL Voluette
Water, deionized ................................................................................................................. 4 liters ....................272-56
IronTot_AVPP_Other_FVR_Eng_Ody.fm
3. Touch
. Touch
On
Adjust
concentration is used, touch the number in the box to enter the actual
concentration, then touch
See Section 3.2.4 Adjusting the Standard Curve on page 40 for more information.
Precision
Standard: 1.000 mg/L Fe
Program
265
267
See Section 3.4.3 Precision on page 44 for more information, or if the standard
concentration did not fall within the specified range.
Sensitivity
Program
Portion of Curve
265
267
See Section 3.4.5 Sensitivity on page 44 for more information.
FerroVer Iron Reagent converts all soluble iron and most insoluble forms of iron
in the sample to soluble ferrous iron. The ferrous iron reacts with the
1,10 phenanthroline indicator in the reagent to form an orange color in
proportion to the iron concentration. Test results are measured at 510 nm.
®
Ampuls ......................................... 1 ampul........... 25/pkg...............25070-25
®
Ampule, 50-mg/L as Fe ............................ 16/pkg...............14254-10
to accept the displayed concentration. If an alternate
. Touch
.
OK
Adjust
95% Confidence Limits of Distribution
0.989–1.011 mg/L Fe
0.977–1.023 mg/L Fe
∆Abs
Entire range
0.010
Entire range
0.010
Quantity Required
per test
Iron, Total
∆Concentration
0.022 mg/L Fe
0.023 mg/L Fe
Unit
Cat. No.
Iron, Total
Page 5 of 6