Bromine
5.
Fill a second sample
cell with 10 mL of
sample. (This is the
blank.)
AccuVac
Hach Programs
1.
Touch
.
Hach Programs
Select program
.
55 Bromine, AV
Touch
.
Start
Bromine
Page 2 of 4
10 mL
6.
Wipe the blank and
place it into the cell
holder.
®
Ampul
2.
Fill a sample cell with
10 mL of sample. (This is
the blank.)
Note: Collect at least 40 mL of
sample in a 50-mL beaker.
Zero
7.
Touch
.
Zero
The display will show:
0.00 mg/L Br
2
3.
Fill a DPD Total
Chlorine Reagent
AccuVac Ampul with
sample. Keep the tip
immersed while the
ampule fills completely.
10 mL
8.
Within three minutes
after the timer beeps,
wipe the prepared
sample and place it into
the cell holder.
Touch
.
Read
Results will appear in
mg/L Br
.
2
Method 8016
4.
Quickly invert the
ampules several times to
mix. Wipe off liquid or
fingerprints.
A pink color will develop
if bromine is present.