DR/2400
UniCell™ Vials
Scope and Application:
surface water, and process control
Tips and Techniques
• Adjust the pH of preserved samples to between pH 2–10 before testing.
• Clean the outside of the vial before placing it in the instrument. Wipe with a damp towel, followed by a dry one.
• Use the standard solution or standard addition method to verify results (see Accuracy Check on page 3).
• See Section 3.2.4 Adjusting the Standard Curve on page 40 for information on adjusting the calibration curve.
• Make sure that the temperature of the water sample and the sample vial is between 18–20 °C. Temperatures outside of this range
will affect the color of the reaction and may give inaccurate results.
• Underrange appears on the display when the instrument is zeroed or when the determined sample concentration is below the
operating range listed for this method.
UniCell Vials
Hach Programs
1.
Touch
.
Hach Programs
Select program
804 Phosphate, HCT 122
Touch
.
Start
PhosphorusReac_UniCell_Other_Asc_Eng_Ody.fm
For water, wastewater, boiler water,
2.
Pipet 0.4 mL of
sample into a sample
vial.
.
Phosphorus, Reactive
(Orthophosphate)
(6.0 to 60.0 mg/L PO
3.
Pipet 0.5 mL
of Reagent B
(HCT 121/122 B) into
the sample vial.
Phosphorus, Reactive (Orthophosphate)
Ascorbic Acid Method
3–
4
4.
Screw a grey
UniCap C (HCT 122 C)
onto the sample vial.
Page 1 of 4
)