DR/2000 and DR/2010 Spectrophotometer
Wavelength Calibration Adjustment
Purpose
These instructions provide a step-by-step procedure for performing a wavelength calibration
adjustment for the DR/2000 Spectrophotometer (software versions 2.0, 2.2, 3.0 and greater) and
DR/2010 Spectrophotometer. Do not perform this procedure if your instrument has 1.261 or 1.27
software! Perform the adjustment any time the lamp assembly (Cat. No. 46644-00) is replaced
and periodically for optimum performance. The adjustment assures the best possible accuracy
and repeatability. Use the Calibration Filter Assembly (Cat. No. 46646-00) to make the
adjustment.
Start by following the three steps below. Then choose the step-by-step instructions that apply to
the specific instrument and software you are using to finish the calibration adjustment.
1. If lamp replacement is necessary, see the Lamp Replacement instructions shipped with
2. Determine which software version applies to the spectrophotometer being adjusted. For the
Select the appropriate wavelength calibration adjustment procedure on the following pages
and continue.
the lamp.
DR/2010 Spectrophotometer, the software version will briefly appear after the instrument is
turned on. For the DR/2000 Spectrophotometer, different messages will appear, depending
on the software in the instrument (see Table 1):
Table 1 DR/2000 Messages Indicating Software Versions
DR/2000 Software Version
2.0 or 2.2
3.0 and greater
1.261 and 1.27
If the first display is
Method #?
then enter 888 with the numeric keys. Press READ/ENTER. The
V2.0
V2.2
display will show
or
Self-Test V3.X
DO NOT perform the calibration adjustment. Monochromator
calibration could be permanently lost (Hach Service cannot restore
the calibration).
Display
, press SHIFT CONFIG/METH,
.
46644-88

Advertisement

loading

Summary of Contents for Hach DR/2000

  • Page 1 Purpose These instructions provide a step-by-step procedure for performing a wavelength calibration adjustment for the DR/2000 Spectrophotometer (software versions 2.0, 2.2, 3.0 and greater) and DR/2010 Spectrophotometer. Do not perform this procedure if your instrument has 1.261 or 1.27 software! Perform the adjustment any time the lamp assembly (Cat. No. 46644-00) is replaced and periodically for optimum performance.
  • Page 2 Wavelength Calibration Adjustment, continued DR/2010 Spectrophotometer Procedure 850 nm Empty the cell Press Scroll to the Adjust nm Press: . The SHIFT SETUP/2 ENTER compartment and close the enter the Setup Menu. Press option with the display will show: UP ARROW cover.
  • Page 3 The display shows . Press twice Completed EXIT to get the Enter Program # prompt. DR/2000 Spectrophotometer Procedure, Version 3.0 and Greater 850 nm Press: Press Press to select With the cell CONFIG/METH SHIFT % T twice to return to the the %T mode.
  • Page 4 READ/ENTER ADJUST nm transmittance. If the wavelength calibration adjustment was successful, the display will show: COMPLETED with 808 in the nm field and then: ADJUST nm If you cannot obtain a good adjustment, contact the Hach Service center serving you.
  • Page 5 Wavelength Calibration Adjustment, continued DR/2000 Spectrophotometer Procedure, Versions 2.0 and 2.2 850 nm Press: Press three Press to select With the cell READ/ENTER SHIFT %T times to accept the mode the %T mode. The display compartment empty and the SHIFT CONFIG/METH...
  • Page 6 METHOD #? prompt. If you cannot obtain a good adjustment, contact the Hach Service center serving you. © Hach Company, 1999. All rights reserved. Printed in the U.S.A. hm/dk 6/99 1ed...