Reagent Blank Adjustment
Accuracy Check
Lead_None_Other_LTK_Eng_Ody.fm
Sampling for Lead Contamination in Household Pipes for Point-of-use
Drinking Water
The sample should be collected after sitting in pipes with no flow for 8 to 18
hours.
Add 10 mL of pPb-1 Acid Preservative (Cat. No. 23685-31) to a one-liter
bottle.
Turn on tap and collect exactly the first liter of water in the bottle containing
acid preservative.
Cap and invert several times to mix.
After two minutes the sample is ready for analysis. Steps 3 and 4 are skipped
in the analysis procedure. Use 100 mL of this preserved sample directly in
step 5.
Sampling for Lead Contamination from Drinking Water Sources Such as Well
Water or Water from Main Supply Lines
Add 10 mL of pPb-1 Acid Preservative (Cat. No. 23685-31) to a one-liter
bottle.
Turn on the tap for 3–5 minutes or until the water temperature has been
stable for 3 minutes.
Collect exactly one liter of water into the bottle containing the acid
preservative.
Cap and invert several times to mix.
After two minutes the sample is ready for analysis. Steps 3 and 4 are skipped
in the analysis procedure. Use 100 mL of this preserved sample directly in
step 5.
At least one liter should be collected to obtain a representative sample. If less
than one liter is collected, use 1 mL of pPb-1 Acid Preservative per 100 mL of
sample.
If nitric acid is to be substituted for pPb-1 as a preservative or the sample is
digested, the buffering capacity of the pPb-2 Fixer Solution
(Cat. No. 23686-55) may be exceeded. Adjust the sample pH to 6.7–7.1 pH
with 5 N Sodium Hydroxide (Cat. No. 2450-53) after step 6.
®
The LeadTrak
program will allow a reagent blank value from -5 to +5 µg/L Pb
to be automatically subtracted from the test result. When using the reagent
blank adjustment feature, the concentration value displayed after zeroing should
be 0 µg/L.
Standard Additions Method (Sample Spike)
1. After reading test results, leave the sample cell (unspiked sample) in the
instrument.
2. Touch
. Touch
Options
additions procedure will appear.
3. Touch
to accept the default values for standard concentration, sample
OK
volume, and spike volumes. Touch
are accepted, the unspiked sample reading will appear in the top row. See
Standard Additions in the instrument manual for more information.
. A summary of the standard
Standard Additions
to change these values. After values
Edit
Lead
Lead