Accuracy Check
Nickel_UniCell_Other_DIMETH_Eng_Ody.fm
Standard Additions Method (Sample Spike)
1. After reading test results, leave the sample cell (unspiked sample) in the
instrument.
2. Touch
. Touch
Options
additions procedure will appear.
3. Touch
to accept the default values for standard concentration, sample
OK
volume, and spike volumes. Touch
are accepted, the unspiked sample reading will appear in the top row. See
Standard Additions in the instrument manual for more information.
4. Prepare three sample spikes. Fill three mixing cylinders with 100 mL of
sample. Use a pipet to add 0.1 mL, 0.2 mL, and 0.3 mL of 1000-mg/L
Ni standard, respectively, to each sample and mix each thoroughly.
5. Transfer 4 mL of each solution into a sample vial and analyze as described in
the procedure. Accept each standard additions reading by touching
Each addition should reflect approximately 100% recovery.
6. After completing the sequence, touch
the standard additions data points, accounting for matrix interferences. Touch
, then select
View: Fit
the sample spikes and the "Ideal Line" of 100% recovery.
See Section 3.2.2 Standard Additions on page 32 for more information.
Standard Solution Method
1. Prepare a 2.00-mg/L Nickel stock solution by pipetting 0.20 mL of Nickel
Standard Solution, 1000-mg/L as Ni, into a 100-mL volumetric flask. Dilute
to the mark with deionized water. Prepare this solution daily.
2. To adjust the calibration curve using the reading obtained with the
2.00-mg/L working solution, touch
Touch
Standard Adjust
3. Touch
. Touch
On
Adjust
concentration is used, touch the number in the box to enter the actual
concentration, then touch
See Section 3.2.4 Adjusting the Standard Curve on page 40 for more information.
. A summary of the standard
Standard Additions
to change these values. After values
Edit
Graph
and touch
Ideal Line
OK
Options
.
to accept the displayed concentration. If an alternate
. Touch
.
OK
Adjust
Nickel
Read
to view the best-fit line through
to view the relationship between
on the current program menu.
Page 3 of 4
.