Hach DR2400 Manual page 338

Hide thumbs Also See for DR2400:
Table of Contents

Advertisement

Iron, Total
5.
Fill another sample
cell (the blank) with
10 mL of sample.
AccuVac Ampul
Hach Programs
1.
Touch
.
Hach Programs
Select program
.
267 Iron, FerroVer AV
Touch
.
Start
Iron, Total
Page 2 of 6
10 mL
6.
When the timer
beeps, place the blank
into the cell holder.
25 mL
20 mL
10 mL
2.
Fill a sample cell with
25 mL of sample. Collect
at least 40 mL of sample
in a 50-mL beaker.
Zero
7.
Touch
.
Zero
The display will show:
0.00 mg/L Fe
3.
Fill a FerroVer Iron
AccuVac® Ampul with
sample.
Keep the tip immersed
while the ampule fills
completely.
10 mL
8.
Place the prepared
sample into the cell
holder.
Touch
.
Read
Results will appear in
mg/L Fe.
Method 8008
4.
Quickly invert the
ampule several times to
mix. Wipe off any liquid
or fingerprints.
IronTot_AVPP_Other_FVR_Eng_Ody.fm