DR/2400
Method 10023
Test 'N Tube™ Vials
Scope and Application:
* Adapted from Clin. Chim. Acta, 14, 403 (1966)
Tips and Techniques
• The ammonia salicylate reagent contains sodium nitroferricyanide. Cyanide solutions are regulated as hazardous wastes by the
Federal RCRA. Collect cyanide solutions for disposal as reactive (D001) waste. Be sure cyanide solutions are stored in a caustic
solution with pH >11 to prevent release of hydrogen cyanide gas. See Section 3 for further information in proper disposal of these
materials.
• Wipe the outside of sample cells before each insertion into the instrument cell holder. Use a damp towel followed by a dry one to
remove fingerprints or other marks.
• A green color will develop if ammonia is present.
Test 'N Tube
Hach Programs
Hach Programs
1.
Touch
.
Hach Programs
Select program
342 N, Ammonia LR TNT
Touch
.
Start
NitrogenAmm_TNT_Low_SAL_Eng_Ody.fm
For water, wastewater, and seawater
2.
Add 2.0 mL of sample
to one AmVer™ Diluent
Reagent Test 'N Tube™
for Low Range Ammonia
Nitrogen (this is the
.
prepared sample).
Nitrogen, Ammonia
LR (0.02 to 2.50 mg/L NH
3.
Add 2.0 mL of
ammonia-free water to
another AmVer™ Diluent
Reagent Test 'N Tube
for Low Range Ammonia
Nitrogen (this is the
blank).
Salicylate Method*
–N)
3
Method 10023
4.
Add the contents of
one Ammonia Salicylate
Reagent Powder Pillow
to each vial.
Nitrogen, Ammonia