Manganese
Method Performance
Summary of Method
Required Reagents
Description
Manganese Reagent Set, 10-mL (50 tests) ....................................................................................................26517-00
Includes:
Alkaline Cyanide Reagent ................................................................ 30 drops.......... 50 mL SCDB .....21223-26
Ascorbic Acid Powder Pillows ........................................................ 2 pillows ......... 100/pkg.............14577-99
PAN Indicator Solution, 0.1%........................................................... 42 drops.......... 50 mL SCDB .....21224-26
Water, deionized ....................................................................................... 10 mL .............. 4 liters ....................272-56
Required Apparatus
Sample Cells, 10-20-25 mL, w/cap......................................................... 2 ....................... 6/pkg.................24019-06
OPTIONAL REAGENTS AND STANDARDS
Manganese Standard Solution, 10-mg/L Mn, 2-mL ampule ....................................... 20/pkg...............26058-20
Manganese Standard Solution, 250-mg/L Mn, 10-mL Voluette
FOR TECHNICAL ASSISTANCE, PRICE INFORMATION AND ORDERING:
In the U.S.A. –
Outside the U.S.A. –
On the Worldwide Web –
© Hach Company, 2003. All rights reserved. Printed in the U.S.A.
Page 4 of 4
Precision
Standard: 0.500 mg/L Mn
Program
290
See Section 3.4.3 Precision on page 44 for more information, or if the standard
concentration did not fall within the specified range.
Sensitivity
Portion of Curve
Entire range
See Section 3.4.5 Sensitivity on page 44 for more information.
The PAN method is a highly sensitive and rapid procedure for detecting low
levels of manganese. An ascorbic acid reagent is used initially to reduce all
oxidized forms of manganese to Mn
mask any potential interferences. PAN Indicator is then added to combine with
2+
the Mn
to form an orange-colored complex. Test results are measured at
560 nm.
Call
toll-free
800-227-4224
Contact the HACH office or distributor serving you.
www.hach.com
; E-mail –
techhelp@hach.com
95% Confidence Limits of Distribution
0.492–0.508 mg/L Mn
∆Abs
0.010
2+
. An alkaline-cyanide reagent is added to
Quantity Required
per test
®
ampule ................. 16/pkg...............14258-10
Manganese_PP_Low_PAN_Eng_Ody.fm
∆Concentration
0.007 mg/L Mn
Unit
Cat. No.
HACH COMPANY