Carrier Waveform Shape; Carrier Frequency - Agilent Technologies 33250A User Manual

80 mhz function / arbitrary waveform generator
Hide thumbs Also See for 33250A:
Table of Contents

Advertisement

3
Chapter 3 Features and Functions
Amplitude Modulation (AM)

Carrier Waveform Shape

• AM carrier shape: Sine, Square, Ramp, or Arbitrary waveform.
The default is Sine. You cannot use pulse, noise, or dc as the
carrier waveform.
• Front-Panel Operation: Press any of the front-panel function keys
except
or
choose the Select Wform softkey to select the active waveform.
• Remote Interface Operation:
FUNCtion:SHAPe {SINusoid|SQUare|RAMP|USER}
You can also use the APPLy command to select the function, frequency,
amplitude, and offset with a single command.

Carrier Frequency

The maximum carrier frequency depends on the function selected as
shown below. The default is 1 kHz for all functions.
Function
Minimum Frequency
1 µHz
Sine
1 µHz
Square
1 µHz
Ramp
1 µHz
Arbs
• Front-Panel Operation: To set the carrier frequency, press the Freq
softkey for the selected function. Then use the knob or numeric
keypad to enter the desired frequency.
• Remote Interface Operation:
FREQuency {<frequency>|MINimum|MAXimum}
You can also use the APPLy command to select the function, frequency,
amplitude, and offset with a single command.
68
. For arbitrary waveforms, press
Maximum Frequency
80 MHz
80 MHz
1 MHz
25 MHz
and then