Handshake Selection (Rs-232) - Agilent Technologies 33250A User Manual

80 mhz function / arbitrary waveform generator
Hide thumbs Also See for 33250A:
Table of Contents

Advertisement

Handshake Selection (RS-232)

You can select one of several handshake (or "flow control") methods to
coordinate the transfer of data between the function generator and your
computer or modem. The method that you select will be determined by
the handshake mode used by your computer or modem.
You can select the handshake mode from the front panel only.
• Select one of the following: None, DTR/DSR (factory setting),
Modem, RTS/CTS, or XON/XOFF.
• None: In this mode, data is sent and received over the interface
without any flow control used. When using this method, use a slower
baud rate (< 9600 baud) and avoid sending more than 128 characters
without stopping or reading a response.
• DTR/DSR: In this mode, the function generator monitors the state of
the DSR (data set ready) line on the RS-232 connector. When the line
goes true, the function generator sends data over the interface.
When the line goes false, the function generator stops sending
information (typically within six characters). The function generator
sets the DTR line false when the input buffer is almost full (about 100
characters) and releases the line when space is available again.
• Modem: This mode uses the DTR/DSR and RTS/CTS lines to control
the flow of data between the function generator and a modem.
When the RS-232 interface is selected, the function generator sets the
DTR line true. The DSR line is set true when the modem is on-line.
The function generator sets the RTS line true when it is ready to
receive data. The modem sets the CTS line true when it is ready to
accept data. The function generator sets the RTS line false when the
input buffer is almost full (about 100 characters) and releases the line
when space is available again.
• RTS/CTS: This mode operates the same as the DTR/DSR mode but
uses the RTS (request to send) and CTS (clear to send) lines on the
RS-232 connector instead. When the CTS line goes true, the function
generator sends data over the interface. When the line goes false,
the function generator stops sending information (typically within six
characters). The function generator sets the RTS line false when the
input buffer is almost full (about 100 characters) and releases the line
when space is available again.
Chapter 3 Features and Functions
Remote Interface Configuration
3
121

Hide quick links:

Advertisement

Table of Contents
loading