Data out of range;
-222
burst period limited by length of burst; value clipped to upper limit
It is not possible to specify a burst period which is too short for the
function generator to output with the specified burst count and
frequency (see below). If the burst period is too short, the function
generator will automatically adjust it as needed to continuously
re-trigger the burst.
Burst Period >
Data out of range;
-222
burst count limited by length of burst; value clipped to lower limit
If the Immediate trigger source is selected (TRIG:SOUR IMM command),
the burst count must be less than the product of the burst period and the
waveform frequency as shown below.
Data out of range;
-222
amplitude; value clipped to ...
This generic message indicates that the waveform amplitude has been
limited to an upper or lower boundary.
-222
Data out of range;
offset; value clipped to ...
This generic message indicates that the offset voltage has been limited to
an upper or lower boundary.
Data out of range;
-222
frequency in burst mode; value clipped to ...
This generic message indicates that the frequency has been limited to an
upper or lower boundary as dictated by the burst period.
-222
Data out of range;
frequency in FM; value clipped to ...
This generic message indicates that the carrier frequency has been
limited to a lower boundary as determined by the FM:DEV command.
The carrier frequency must always be greater than or equal to the
frequency deviation.
Burst Count
---------------------------------------------------------- -
Waveform Frequency
Burst Count < Burst Period X Waveform Frequency
Chapter 5 Error Messages
Execution Errors
+ 200 ns
4
5
263