Agilent Technologies 33250A User Manual page 193

80 mhz function / arbitrary waveform generator
Hide thumbs Also See for 33250A:
Table of Contents

Advertisement

• When the Bus (software) source is selected, the function generator
outputs one burst each time a bus trigger command is received.
To trigger the function generator from the remote interface
(either GPIB or RS-232), send the TRIG or *TRG (trigger) command.
The front-panel
is waiting for a bus trigger.
• When the External or Bus trigger source is selected, the burst count
and burst phase remain in effect but the burst period is ignored.
• The APPLy command automatically sets the trigger source to
Immediate (equivalent to TRIG:SOUR IMM command).
• To ensure synchronization when the Bus source is selected, send the
*WAI (wait) command. When the *WAI command is executed, the
function generator waits for all pending operations to complete before
executing any additional commands. For example, the following
command string guarantees that the first trigger is accepted and the
operation is executed before the second trigger is recognized.
TRIG:SOUR BUS;*TRG;*WAI;*TRG;*WAI
• You can use the *OPC? (operation complete query) command or the
*OPC (operation complete) command to signal when the burst is
complete. The *OPC? command returns "1" to the output buffer when
the burst is complete. The *OPC command sets the "Operation Complete"
bit (bit 0) in the Standard Event register when the burst is complete.
TRIGger:DELay {<seconds>|MINimum|MAXimum}
TRIGger:DELay? [MINimum|MAXimum]
Insert a time delay between the receipt of the trigger and the start of the
burst waveform (used in triggered burst mode only). Select from 0 seconds
to 85 seconds. The default delay is 0. MIN = 0 seconds. MAX = 85 seconds.
The :DEL? query returns the trigger delay in seconds.
TRIGger:SLOPe {POSitive|NEGative}
TRIGger:SLOPe?
Select whether the function generator uses the rising edge or falling
edge of the trigger signal on the rear-panel Trig In connector for an
externally-triggered burst. The default is POS (rising edge). The :SLOP?
query returns "POS" or "NEG".
Chapter 4 Remote Interface Reference
key is illuminated when the function generator
Burst Mode Commands
193
4

Hide quick links:

Advertisement

Table of Contents
loading