Modulating Waveform Shape; Modulating Waveform Frequency - Agilent Technologies 33250A User Manual

80 mhz function / arbitrary waveform generator
Hide thumbs Also See for 33250A:
Table of Contents

Advertisement

Modulating Waveform Shape

The function generator will accept an internal or external modulation
source for AM.
• Modulating waveform shape (internal source): Sine, Square, Ramp,
Negative Ramp, Triangle, Noise, or Arb waveform. The default is Sine.
• Square has 50% duty cycle.
• Ramp has 100% symmetry.
• Triangle has 50% symmetry.
• Negative ramp has 0% symmetry.
• You can use noise as the modulating waveshape, but you cannot use
noise, pulse, or dc as the carrier waveform.
• If you select an arbitrary waveform as the modulating waveshape,
the waveform is automatically limited to 8K points. Extra waveform
points are removed using decimation.
• Front-Panel Operation: After enabling AM, press the Shape softkey.
• Remote Interface Operation:
AM:INTernal:FUNCtion {SINusoid|SQUare|RAMP|NRAMp|
TRIangle|NOISe|USER}

Modulating Waveform Frequency

The function generator will accept an internal or external modulation
source for AM.
• Modulating frequency (internal source): 2 mHz to 20 kHz.
The default is 100 Hz.
• Front-Panel Operation: After enabling AM, press the AM Freq softkey.
• Remote Interface Operation:
AM:INTernal:FREQuency {<frequency>|MINimum|MAXimum}
Chapter 3 Features and Functions
Amplitude Modulation (AM)
3
69

Hide quick links:

Advertisement

Table of Contents
loading