Agilent Technologies 33250A User Manual page 257

80 mhz function / arbitrary waveform generator
Hide thumbs Also See for 33250A:
Table of Contents

Advertisement

Settings conflict;
-221
not able to sweep dc, modulation turned off
The function generator cannot generate a sweep using the dc voltage
function. The sweep mode has been turned off.
-221
Settings conflict;
not able to burst dc, burst turned off
The function generator cannot generate a burst using the dc voltage
function. The burst mode has been turned off.
-221
Settings conflict;
not able to sweep noise, sweep turned off
The function generator cannot generate a sweep using the noise function.
The sweep mode has been turned off.
-221
Settings conflict;
pulse width decreased due to period
For a pulse waveform, the function generator will automatically adjust
the waveform parameters in the following order as needed to generate
a valid pulse: (1) edge time, (2) pulse width, and then (3) period.
In this case, the function generator has decreased the pulse width to
accommodate the specified period (the edge time is already at its
minimum setting).
-221
Settings conflict;
edge time decreased due to period
For a pulse waveform, the function generator will automatically adjust
the waveform parameters in the following order as needed to generate a
valid pulse: (1) edge time, (2) pulse width, and then (3) period.
In this case, the function generator has decreased the edge time to
accommodate the specified period and preserve the pulse width setting.
Chapter 5 Error Messages
Execution Errors
4
5
257