Agilent Technologies 33250A User Manual page 217

80 mhz function / arbitrary waveform generator
Hide thumbs Also See for 33250A:
Table of Contents

Advertisement

Chapter 4 Remote Interface Reference
System-Related Commands
*OPC
Set the "Operation Complete" bit (bit 0) in the Standard Event register
after the previous commands have been executed. Other commands may
be executed before the bit is set. This command is used in the triggered
sweep or triggered burst modes to provide a way to poll or interrupt the
computer when the sweep or burst is complete.
*OPC?
Return "1" to the output buffer after the previous commands have been
executed. Other commands cannot be executed until this command
completes. Used in the triggered sweep and triggered burst modes only.
*WAI
Wait for all pending operations to complete before executing any additional
commands over the interface. Used in the triggered sweep and triggered
burst modes only.
4
217