Chapter 5 Error Messages
Execution Errors
Queue overflow
The error queue is full because more than 20 errors have occurred.
No additional errors are stored until you remove errors from the queue.
The error queue is cleared by the *CLS (clear status) command or when
the power is cycled. The errors are also cleared when you read the queue.
The error queue is not cleared by an instrument reset (*RST command).
Parity error in program message
This error is most likely to occur when the function generator's parity
setting (RS-232 interface) does not match the computer's parity setting.
This error may also occur in the presence of noise on the RS-232 cable.
Framing error in program message
This error is most likely to occur when the number of stop bits for the
computer (RS-232 interface) does not match the function generator's
setting (fixed at 1 stop bit).
Input buffer overrun
Too many characters have been sent to the function generator while
configured to use the RS-232 interface. This error typically occurs when
you have selected no handshaking of data between the computer and the
function generator. To avoid this error, select one of the handshake
modes available for the 33250A (see "Remote Interface Configuration"
on page 118 for more information).